| CONTACT FUNCTION
The load guard with one or two microswitches is supplied with alternative
contact connections for each microswitch. Load guards with three, four
or five microswitches have either one breaking or closing function on
each switch. These are normally set for the breaking function (closed
contacts at rest) and for rising loads, unless otherwise ordered. When
setting for falling load, e.g. control on the slack line, the microswitch
is connected for the closing function (open contact at rest).
If the load guard is to be used for safety application, the breaking
function should always be used.
The microswitches have gold-plated contacts, which make them suitable
for operation on low current/voltage (under 10 V and/or 60 mA).
The difference between the switching value on rising and falling loads
is approximately 3% of the max. capacity of the instrument. This hysteresis
is somewhat reduced at lower loads and increases slightly at higher load
values. To avoid "chattering" of the contacts and contactors if the load
should start to sway, a time relay to delay the reclosing of the hoisting
circuit may be installed. When setting for falling loads, special microswitches
with a reduced hysteresis are used.
The switch values are set by the factory to the required level.
|